Key result
Riata 1580 ICD leads had a significantly higher risk of failure compared to Endotak 0158 leads over a 10-year follow-up (15.9% vs 6.3%, P=0.0001).
Why the study?
The long-term performance of the Riata family of leads has come under scrutiny, prompting an evaluation of Riata 1580 leads compared with Endotak 0158 leads.
Does the Riata 1580 ICD lead have a higher risk of long-term failure compared to the Endotak 0158 ICD lead?
Cohort (n=689)
No
Does the Riata 1580 ICD lead have a higher risk of long-term failure compared to the Endotak 0158 ICD lead?
Riata 1580 ICD leads demonstrate a significantly higher long-term failure rate compared to Endotak 0158 leads, particularly after 6 years, highlighting the need for careful ongoing surveillance.
Answers are grounded in this paper and cite their sources.
Supports intensified surveillance for Riata 1580 leads; extends failure observations but leaves management strategies open.
BACKGROUND: The long-term performance of the Riata family of leads has recently come under increasing scrutiny. We aimed to determine the long-term performance of the Riata 1580 leads compared with Endotak 0158 leads. METHODS: All patients with Riata 1580 or Endotak 0158 leads implanted from 2003 to 2008 at the Heart Hospital, UCLH were analyzed. Significant electrical changes were as follows: threshold increase >1 V at a set pulse width between pacing checks, persistent R wave fall to <2 mV or reduction in R wave >50%, noise, pacing impedance change to <300 Ω or >1500 Ω, high voltage (HV) change to <20 Ω or >200 Ω, HV change ± 15 Ω, pacing impedance change >400 Ω over 12 months. RESULTS: 333 Riata and 356 Endotak leads were implanted. Median follow-up time + interquartile range were calculated, after exclusion of censored events including loss to follow-up: Riata 3652 + 655 days, Endotak 3730 + 810 days. A total of 51 (15.9%) Riata leads and 21 (6.3%) Endotak leads were affected. A greater risk of failure was found for the Riata lead compared with the Endotak lead (P = .0001). An additional time-dependent effect was found, with the Riata lead 1.9 times more likely to fail in the first 6 years following lead implantation and 5.3 times more likely to fail after 6 years. CONCLUSIONS: Riata leads have a higher risk of failure compared to Endotak leads over time. The importance of careful ongoing performance surveillance late in the leads' lifetime is reflected in this 10-year follow-up study.
No takes yet. Share an insight, caveat, or question.
Cobb et al. (2020) conducted a cohort in ICD lead implantation (n=689). Riata 1580 ICD lead vs. Endotak 0158 ICD lead was evaluated on Significant electrical changes (lead failure) (p=0.0001). Riata 1580 ICD leads had a significantly higher risk of failure compared to Endotak 0158 leads over a 10-year follow-up (15.9% vs 6.3%, P=0.0001).