Key result
Conductors' externalization occurred in 33% of patients with Riata leads and was independently associated with higher LVEF and decreased ventricular electrogram amplitude.
Why the study?
What are the predictors and prevalence of conductor externalization in patients with small-caliber Riata ICD leads?
Observational (n=36)
What are the predictors and prevalence of conductor externalization in patients with small-caliber Riata ICD leads?
In patients with Riata ICD leads, higher LVEF and a progressive decrease in ventricular electrogram amplitude are associated with conductor externalization, which is often preceded by electrical dysfunction.
Answers are grounded in this paper and cite their sources.
May guide targeted monitoring of Riata leads with higher LVEF or declining electrograms; leaves open prospective validation of clinical utility.
AIMS: Increased rates of structural abnormalities including externalized conductors have been reported in the Riata family of implantable cardioverter-defibrillator leads (St Jude Medical). Little is known about their reliability and the time lag for emergence of functional or structural abnormalities. METHODS AND RESULTS: Thirty-six patients who received small-caliber leads of the Riata family and who completed face-profile flouroscopies, repeated at every 6 months were included. We assessed the prevalence of conductors' externalization and its relation to abnormal electrical parameters or adverse events. Thirty-six patients, mean age = 64 ± 10 years, with at least 7-month completed fluoroscopy follow-up were included in the analysis. Externalized conductors were identified in 12 (33%) patients after a 53-month (13-114) mean delay. A higher left ventricular ejection fraction (LVEF): 47 ± 13 vs. 33 ± 12%, P = 0.04, and a progressive decrease (≥30% of the initial value) in amplitude of ventricular electrogram 9/12 (75%) vs. 4/24 (17%), P = 0.03 were independently associated with the fluoroscopic failures. Detection of the conductors' externalization was preceded by an electrical lead abnormality in 10 (83%) patients. CONCLUSION: Repeated face-profile fluoroscopies allowed detection of conductors' externalization in 33% of patients implanted with Riata leads. Better LVEF and a progressive decrease in amplitude of intracardiac ventricular electrogram were independently associated with externalized conductors. The structural abnormality was preceded by an electrical lead dysfunction 83% of patients.
No takes yet. Share an insight, caveat, or question.
Kubala et al. (2013) conducted an observational in Implantable cardioverter-defibrillator (ICD) leads (Riata family) (n=36). Riata family implantable cardioverter-defibrillator leads was evaluated on Conductors' externalization. Conductors' externalization occurred in 33% of patients with Riata leads and was independently associated with higher LVEF and decreased ventricular electrogram amplitude.