Key result
Externalized conductors were found in 14.3% of Riata ICD leads via fluoroscopy, occurring more frequently in 8-F Riata than 7-F Riata ST leads.
Why the study?
What is the prevalence of externalized conductors and electrical abnormalities in Riata ICD leads assessed by fluoroscopic screening?
Cross-Sectional (n=1,029)
Yes
What is the prevalence of externalized conductors and electrical abnormalities in Riata ICD leads assessed by fluoroscopic screening?
Fluoroscopic screening reveals a high prevalence (14.3%) of externalized conductors in Riata ICD leads, the majority of which are not detectable by standard electrical interrogation.
Answers are grounded in this paper and cite their sources.
May warrant fluoroscopic screening for Riata leads; leaves open optimal surveillance and outcome impact.
BACKGROUND: The Riata family of implantable cardioverter-defibrillator (ICD) leads is prone to a specific insulation abrasion characterized by externalization of conductor cables. The objective of this study was to determine the prevalence of externalized conductors and electrical abnormalities in Riata ICD leads by fluoroscopic screening and standard ICD interrogation. METHODS AND RESULTS: All ICD implantation centers were contacted by the Netherlands Heart Rhythm Association Device Advisory Committee to identify all patients with an active Riata ICD lead and to perform fluoroscopic screening of the lead. In addition, the electrical integrity of the lead was assessed. As of March 1, 2012, data for 1029 active Riata leads were available; 47% of these were 8-F Riata and 53% were 7-F Riata ST. Externalized conductors were observed in 147 leads (14.3%). Proportion of externalized conductors was higher in 8-F Riata compared with 7-F Riata ST (21.4% vs 8.0%; P<0.001). Median time from implantation to detection of externalized conductors was 65.3 months. The estimated rates of externalized conductors were 6.9% and 36.6% at 5 and 8 years after implantation, respectively. Of the 147 leads with externalized conductors, 10.9% had abnormal electrical parameters vs 3.5% in nonexternalized leads (P<0.001). CONCLUSIONS: The prevalence of externalized conductors in Riata leads is significantly high (14.3%) using fluoroscopic screening. The majority of externalized conductors are not detectable with standard ICD interrogation. Screening with fluoroscopy is reasonable.
No takes yet. Share an insight, caveat, or question.
Theuns et al. (2012) conducted a cross-sectional in Active Riata ICD lead (n=1,029). 8-F Riata ICD lead vs. 7-F Riata ST ICD lead was evaluated on Externalized conductors (p=<0.001). Externalized conductors were found in 14.3% of Riata ICD leads via fluoroscopy, occurring more frequently in 8-F Riata than 7-F Riata ST leads.