Key result
The Medtronic Sprint Fidelis ICD lead had decreased survival compared with the Quattro lead, with patients under 50 at increased risk of fracture.
Why the study?
Does the Fidelis ICD lead have a higher fracture rate compared to the Quattro lead, and what are the risk factors for fracture?
Observational (n=1,314)
Does the Fidelis ICD lead have a higher fracture rate compared to the Quattro lead, and what are the risk factors for fracture?
Answers are grounded in this paper and cite their sources.
INTRODUCTION: The Medtronic Sprint Fidelis implantable cardioverter defibrillator (ICD) lead was "recalled" in October 2007 after 268,000 implants worldwide due to increased failure risk. Manufacturer suggested monitoring has not been shown effective at preventing adverse events. Only limited data exist regarding clinical predictors of Fidelis lead fracture. We sought to identify risk factors for Fidelis fracture to guide clinical monitoring and compare its performance with a control lead. METHODS: Fractured lead cases were retrospectively reviewed for demographic data, implant technique, radiographic appearance and clinical presentation was analyzed. Lead survival was compared using Kaplan-Meir curves. RESULTS: Study patients (n = 1314) experienced 18 Fidelis and 6 Quattro lead fractures. Patients with failed Fidelis leads were younger than those with surviving leads (49.5 vs 64.6 years, P = 0.0066). Fidelis lead fractures often occurred around the time of physical activity. No other measured demographic or technique related factors were associated with lead fracture. Fidelis leads had significantly decreased survival compared with Quattro leads (89.3 vs 98.9% at 30 months). Patients less than 50 years old had significantly decreased lead survival compared with those older than 50 in both Fidelis (79.6% vs 96.5% at 24 months) and Quattro (93.4 vs 99.8%, P < 0.001 at 24 months) leads. CONCLUSIONS: Patients under age 50, with either Fidelis or Quattro ICD leads, are at increased risk of lead fracture compared with patients over 50, particularly around the time of intense physical activity. Aggressive monitoring and advisory programming appears warranted in patients with Fidelis leads as well as especially in younger patients.
No takes yet. Share an insight, caveat, or question.
Morrison et al. (2010) conducted an observational in Implantable cardioverter defibrillator (ICD) leads (n=1,314). Medtronic Sprint Fidelis ICD lead vs. Quattro ICD lead was evaluated on Lead survival at 30 months. The Medtronic Sprint Fidelis ICD lead had decreased survival compared with the Quattro lead, with patients under 50 at increased risk of fracture.
Synapse has enriched 10 related papers. Consider them for comparative context: