Key result
Reported ICD lead survival varies significantly across studies, ranging from 91-99% at 2 years to 60-72% at 8 years, highlighting the need for a comprehensive registry to monitor lead performance.
Due to highly variable reported survival rates of ICD leads and small existing cohorts, there is a critical need for a comprehensive Europe-wide registry to accurately monitor long-term lead performance and safety.
Answers are grounded in this paper and cite their sources.
Variable ICD lead survival data warrant caution in device counseling; leaves open true long-term failure rates pending a unified registry.
Although the use of implantable cardioverter defibrillator (ICD) therapy has increased over the last decades, the reporting of ICD lead performance is inadequate. So far, there is neither a European nor worldwide registry on ICD leads. The published long-term results from national or multicentre registries encompass relatively small patient cohorts. Nevertheless, the failure of ICD leads may have substantial clinical consequences, including failure to sense, failure to pace, failure to defibrillate, inappropriate shocks, and even death of the patient. The reported ICD lead survival varies significantly between studies: 91-99% at 2 years, 85-95% at 5 years, and 60-72% at 8 years. Thus, the true incidence of lead malfunction cannot be defined as outlined in the present review. One current initiative of the European Heart Rhythm Association is to initiate and develop a Europe-wide registry to monitor, over a prolonged follow-up period, the performance of ICDs and ICD leads.
No takes yet. Share an insight, caveat, or question.
Goette et al. (2008) conducted a review in Implantable cardioverter defibrillator (ICD) lead performance. Implantable cardioverter defibrillator (ICD) leads was evaluated on ICD lead survival. Reported ICD lead survival varies significantly across studies, ranging from 91-99% at 2 years to 60-72% at 8 years, highlighting the need for a comprehensive registry to monitor lead performance.
Synapse has enriched 10 related papers. Consider them for comparative context: