Key result
Capacitance transducers were developed to measure shortening up to 2.5 cm and tension up to 100 g in isolated muscle.
The developed capacitance transducers allow for precise measurement of shortening and tension in isolated muscle preparations.
Answers are grounded in this paper and cite their sources.
Enables precise ex vivo muscle mechanics studies; leaves open clinical translation pending in vivo validation.
Apparatus has been developed for the measurement of shortening and tension of isolated muscle using variable capacitance methods. An isotonic transducer is described which will measure up to 2.5 cm of shortening with a minimum of inertial loading with 1% linearity. An isometric transducer is described which will measure tensions up to 100 g with not more than 5-μ displacement from 0 to 200 cycles/sec. Simple modifications provide other ranges of tension.
No takes yet. Share an insight, caveat, or question.
Myron O. Schilling (1960) studied this question. Capacitance transducers was evaluated on Measurement of shortening and tension of isolated muscle. Capacitance transducers were developed to measure shortening up to 2.5 cm and tension up to 100 g in isolated muscle.