A comparison of retrieved total column ozone (TCO) amounts between the Pandora #34 spectrometer system and the Dobson #061 spectrophotometer from direct-sun observations was performed on the roof of the Boulder, Colorado, NOAA building. This paper, part of an ongoing study, covers a 1-year period starting on 17 December 2013. Both the standard Dobson and Pandora TCO retrievals required a correction, TCOcorr = TCO (1 + C(T)), using a monthly varying effective ozone temperature, TE, derived from a temperature and ozone profile climatology. The correction is used to remove a seasonal difference caused by using a fixed temperature in each retrieval algorithm. The respective corrections C(TE) are CPandora = 0.00333(TE-225) and CDobson = -0.0013(TE-226.7) per degree K. After the applied corrections removed most of the seasonal retrieval dependence on ozone temperature, TCO agreement between the instruments was within 1 % for clear-sky conditions. For clear-sky observations, both co-located instruments tracked the day-to-day variation in total column ozone amounts with a correlation of r2 = 0.97 and an average offset of 1.1 ± 5.8 DU. In addition, the Pandora TCO data showed 0.3 % annual average agreement with satellite overpass data from AURA/OMI (Ozone Monitoring Instrument) and 1 % annual average offset with Suomi-NPP/OMPS (Suomi National Polar-orbiting Partnership, the nadir viewing portion of the Ozone Mapper Profiler Suite).
No takes yet. Share an insight, caveat, or question.
Herman et al. (2015) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: