It is designed to prevent cerebrospinal fluid (CSF) leakage after intradural surgery. This study assessed the safety and biodegradability of Liqoseal in a porcine craniotomy model. Methods: In 32 pigs a craniotomy plus durotomy was performed. In 15 pigs Liqoseal was implanted, in 11 control pigs no sealant was implanted and in 6 control pigs a control dural sealant (Duraseal or Tachosil) was implanted. The safety of Liqoseal was evaluated by clinical, MRI and histological assessment. The degradation of Liqoseal was histologically estimated. Results: Liqoseal, 2 mm thick before application, did not swell and significantly was at maximum mean thickness of 2.14 (±0.37) mm at one month. The foreign body reaction induced by Liqoseal, Duraseal and Tachosil were comparable. Liqoseal showed no adherence to the arachnoid layer and was completely resorbed between 6 and 12 months postoperatively. In one animal with Liqoseal, an epidural fluid collection containing CSF could not be excluded. Conclusion: Liqoseal seems to be safe for intracranial use and is biodegradable. The safety and performance in humans needs to be further assessed in clinical trials.