Key result
The fully automated seismocardiogram annotation procedure correctly detected 91.4% of heartbeats with minimal false positives and showed high linearity (R2 > 0.95) compared to ECG.
Why the study?
Does a fully automated annotation procedure for seismocardiogram traces accurately detect heartbeats compared to electrocardiogram?
Does a fully automated annotation procedure for seismocardiogram traces accurately detect heartbeats compared to electrocardiogram?
An automated seismocardiogram annotation algorithm shows high accuracy and correlation with ECG, enabling reliable, contactless monitoring of heart-related parameters.
May enable contactless cardiac monitoring; leaves open prospective clinical validation.
This paper presents a fully automated procedure for acquiring and analyzing seismocardiographic (SCG) traces from an inertial measurement unit (IMU) placed over a subject's sternum. An automated calibration procedure allows for straightforward adaption to different subjects. Calibration is performed once per subject, exploiting electrocardiogram (ECG) markers; relevant patterns and parameters are automatically extracted and used for successive SCG processing, which does not require concurrent ECG information any longer. Annotation of SCG traces is performed in two steps: in the first one, a suitably engineered signal is derived from SCG and used as coarse heartbeat detector; then, the annotation can be performed by comparing the prototype extracted at calibration time with segments of SCG data, near to the detected beats. The proposed methodology is validated by direct comparison with ECG, adopted as gold standard. In particular, three main metrics are taken into account: sensitivity (i.e., the percentage of correctly identified heartbeats, compared to ECG), precision (i.e., impact of false positives on truly detected beats), and R2(i.e., linearity between beat-to-beat measurements as computed by ECG and SCG). Results show satisfactory performance, more than adequate to continuous, long-term monitoring: overall, approximately 90% of heartbeats are correctly detected, on average, with minimal false positives (≈ 1%). Linearity between ECG- and SCG-computed beat-to-beat intervals is extremely high (R2> 0.95, on average), indicating good agreement between the two measurement methods. These results suggest that SCG can be used as a reliable, contactless measure of heart-related parameters.
No takes yet. Share an insight, caveat, or question.
Mora et al. (2019) studied Healthy (n=15). Automated seismocardiogram (SCG) annotation vs. Electrocardiogram (ECG) was evaluated on Sensitivity of heartbeat detection (ALL condition). The fully automated seismocardiogram annotation procedure correctly detected 91.4% of heartbeats with minimal false positives and showed high linearity (R2 > 0.95) compared to ECG.