Key result
A motion artifact removal scheme exploiting tunable-Q factor wavelet transform domain sparsity achieved a minimum mean absolute error of 1.03 and 0.76 bpm for heart rate estimation.
Why the study?
Classical filters fail to remove motion artifacts covering wide signal frequencies in wrist PPG signals acquired during physical activity.
A novel TQWT and BPDN-based denoising scheme effectively removes motion artifacts from wrist PPG signals without needing reference accelerometer data, achieving highly accurate heart rate estimation.
Wearable health monitoring devices based on photoplethysmogram (PPG) can estimate valuable physiological parameters and are often regarded as a touchstone for indicating the cardiovascular status of a person. Severe motion artifacts are generally observed in the acquired PPG signals from the biosensors in the wearables when the PPG acquisition is performed during different physical activities. Classical filters can remove thermal noise and electromagnetic interference from the raw PPG signals, but they fail to handle the motion artifacts that cover a wider range of signal frequencies. This work introduces a new motion artifact removal scheme that exploits the sparsity of PPG signals in the tunable-Q factor wavelet transform (TQWT) domain using the basis pursuit denoising (BPDN) scheme. The proposed scheme deftly removes noise while preserving the original morphology of the PPG signal without using any reference accelerometer sensor data. In particular, a regularized SALSA-based BPDN algorithm is employed for sparse recovery, which utilizes a soft thresholding approach to denoise the noisy PPG signals. TQWT can be tuned according to the oscillatory behavior of the raw PPG signal. At the same time, BPDN exploits the sparsity of the wavelet coefficients to yield an artifact-free reconstruction of PPG. Further, heart rate estimation is performed using the denoised PPG signal with support vector regression to demonstrate the clinical information preservation during the proposed denoising process. The proposed denoising scheme is evaluated on two publicly available databases, and an overall minimum mean absolute error (MAE) of 1.03 and 0.76 beats per minute (bpm) is achieved. The suggested denoising approach has demonstrated better performance with the lowest MAE compared to the previously reported methods.
No takes yet. Share an insight, caveat, or question.
Gupta et al. (2023) studied this question. Tunable-Q factor wavelet transform (TQWT) domain sparsity using basis pursuit denoising (BPDN) vs. Previously reported methods was evaluated on Heart rate estimation mean absolute error (MAE). A motion artifact removal scheme exploiting tunable-Q factor wavelet transform domain sparsity achieved a minimum mean absolute error of 1.03 and 0.76 bpm for heart rate estimation.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: