Validates an accurate method for detecting chlorate in dairy rinse water, suggesting improved safety monitoring.
Ensuring dairy product safety and quality requires effective self-monitoring tools compliant with regulatory standards. However, no analytical method currently meets the ASSURED criteria for monitoring chlorate (ClO3−) in rinse water or milk within the dairy industry. This study reports the development and validation of a rapid spectrophotometric method for ClO3− quantification designed for on-site self-monitoring. The assay is based on the oxidation of indigo carmine (IC) in a microplate format, where the decrease in absorbance is proportional to the ClO3− concentration, enabling straightforward quantification. Method validation using the method accuracy profile (MAP) approach demonstrated satisfactory trueness and precision. The method achieved a limit of detection (LOD) of 70.75 μg/L and a limit of quantification (LOQ) of 233.5 μg/L in tap water, both well below the regulatory limit of 700 μg/L. Selectivity tests confirmed limited interference, mainly from chlorite and sodium hypochlorite. Overall, the proposed method provides a simple, robust, and cost-effective alternative for rapid screening of ClO3− contamination in dairy industrial settings, supporting improved process control and food safety.
No takes yet. Share an insight, caveat, or question.
Ali et al. (2026) studied this question.