BACKGROUND AND PURPOSE: In this paper, we investigate the possibility to use X-ray based real time 2D/3D registration for non-invasive tumor motion monitoring during radiotherapy. MATERIALS AND METHODS: The 2D/3D registration scheme is implemented using general purpose computation on graphics hardware (GPGPU) programming techniques and several algorithmic refinements in the registration process. Validation is conducted off-line using a phantom and five clinical patient data sets. The registration is performed on a region of interest (ROI) centered around the planned target volume (PTV). RESULTS: The phantom motion is measured with an rms error of 2.56 mm. For the patient data sets, a sinusoidal movement that clearly correlates to the breathing cycle is shown. Videos show a good match between X-ray and digitally reconstructed radiographs (DRR) displacement. Mean registration time is 0.5 s. CONCLUSIONS: We have demonstrated that real-time organ motion monitoring using image based markerless registration is feasible.
No takes yet. Share an insight, caveat, or question.
Gendrin et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: