Key result
Ultrasound-controlled paclitaxel release balloons significantly improved drug content in target porcine coronary arteries to 203 ± 37 μg/g compared to 85 ± 23 μg/g without ultrasound.
Why the study?
Low availability of the drug coating remains a major concern of current DCBs.
Does an ultrasound-controlled paclitaxel release balloon improve drug content in target vessels in porcine coronary arteries?
Does an ultrasound-controlled paclitaxel release balloon improve drug content in target vessels in porcine coronary arteries?
An ultrasound-controlled paclitaxel release balloon significantly improves drug bioavailability in target vessels in a porcine model.
OBJECTIVES: The aim of this study is to improve local-drug delivery efficiency and tissue absorption using the ultrasound (US)-responsible drug coating based on a newly developed US-controlled paclitaxel release balloon. BACKGROUND: Low availability of the drug coating remains a major concern of the current drug coated balloon (DCB). The goal of this study is to develop a method to use an US-responsible paclitaxel-loaded microcapsules (PM) as the main content of balloon drug coating to enhance bioavailability of DCB. METHODS: An US-controlled paclitaxel release balloon is designed and fabricated based on the US-responsible paclitaxel-loaded poly (lactic-co-glycolic acid) (PLGA) microcapsules. Rapid exchange percutaneous transluminal coronary angioplasty (PTCA) balloon catheters were coated with the PM. The deployment processes of the paclitaxel-loaded microcapsules coated balloons (PMCB) under US, PMCB without US and a homogenous matrix of paclitaxel and iopromide coated balloon (PICB) were then placed in healthy and stent implanted porcine coronary arteries. The paclitaxel content in target vessels are 203 ± 37 μg/g for PMCB under US, 85 ± 23 μg/g for PMCB without US, and 107 ± 31 μg/g for PICB 1-hr post-surgery. The availability of the drug for the PMCB reaches 27% under US. CONCLUSIONS: The US-controlled paclitaxel release balloon significantly improved the drug content of the target vessels in the porcine model.
No takes yet. Share an insight, caveat, or question.
Song et al. (2019) studied Coronary artery disease (porcine model). Ultrasound-controlled paclitaxel-loaded microcapsules coated balloons (PMCB under US) vs. PMCB without US and paclitaxel/iopromide coated balloon (PICB) was evaluated on Paclitaxel content in target vessels 1-hr post-surgery. Ultrasound-controlled paclitaxel release balloons significantly improved drug content in target porcine coronary arteries to 203 ± 37 μg/g compared to 85 ± 23 μg/g without ultrasound.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: