Answers are grounded in this paper and cite their sources.
Screw-shaped implants with 3 different surface topographies, evidenced visually as well as numerically with an optical profilometer, were inserted in rabbit bone. After a healing period of 12 weeks, a statistically significant higher removal torque was needed to unscrew screws blasted with 25-microns TiO2 particles and screws blasted with 75-microns particles of AI2O3 compared with screws with a turned surface. The histomorphometric evaluation demonstrated a higher percentage of bone-to-metal contact for implants blasted with 25-microns particles of TiO2 compared with the as-machined implants. A greater surface area of bone in threads was found for the turned screws compared with screws blasted with 25-microns TiO2 particles. In the short-term follow-up, there was a better fixation of implants with an average surface roughness of 0.9-1.3-microns and with a homogeneous surface structure than of implants with an average surface roughness of 0.4 microns and with a clear direction of the surface pattern.
No takes yet. Share an insight, caveat, or question.
Wennerberg et al. (1995) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: