Answers are grounded in this paper and cite their sources.
Laboratory evaluation demonstrates high analytical precision of a liquid-chromatographic hemoglobin A1c assay, suggesting reduced reliance on oral glucose tolerance tests.
A chromatographic method for determining glycated hemoglobin (Hb A1c) by use of a new monodisperse cation-exchanger has been investigated. Hb A1c was separated from other "minor hemoglobins": Hb F, Hb A3 (the glutathione adduct), and the acetaldehyde adduct in alcoholics. The method was fully automated and a single column could be used for more than 1000 runs. The normal reference interval was 4.0-5.2%; the interval for diabetic outpatients was 5.6-12.4%. Within-run and the between-run CVs were less than 0.9% and 1.7%, respectively. Carbamylation in uremic patients who were undergoing hemodialysis increased the proportion of Hb A1c to 1%. Hb A1c results were compared with results from glucose tolerance tests. In our study, Hb A1c less than 5.5% excluded diabetes: subjects with Hb A1c greater than 6.2% showed diabetes. If blood sampled during fasting had been screened with determinations of glucose and Hb A1c, only 20% of referred subjects would have needed an oral glucose tolerance test for diagnosis of diabetes.
No takes yet. Share an insight, caveat, or question.
Jeppsson et al. (1986) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: