Investigates a new portable imaging system for improved diagnostic capabilities, indicating potential breakthroughs for disabled patients.
Medical imaging devices that combine both the functional and anatomical modalities have proved to be highly accurate in diagnoses. Nonetheless, current SPECT-CT systems are bulky, very costly, and non-portable, restricting their use for disabled patients. A comparison was carried out using GATE/GEANT4 Monte Carlo simulation for scintillator efficiency. The experimental work used a 4×4 pixel array GAGG(Ce)+SiPM detector with SIB616/SIB1632 readout electronics and IQSP580 DAQ system. Hexagonal and pinhole collimators were thoroughly investigated. Two new algorithms for image fusion were created; these include Dynamic Alpha Blending and US-Protected Adaptive Overlay. The results show the superiority of GAGG(Ce) with 3.196 mm spatial resolution and 4.9% energy resolution, better than LYSO (3.391mm, 8.5%). Images with gamma photons were successfully acquired using phantom Tc-99m breast lesions from distances of 5 to 20 cm. Both dynamic algorithms yielded medically useful images with Method 2 having anatomical information and Method 1 with functional details. The US-SPECT device designed provides a major breakthrough in portable hybrid medical imaging, providing bedside capability for the disabled patients as well as lower radiation doses as compared to CT-based hybrid devices.