Key result
Fixed-cycle fetal speckle tracking echocardiography demonstrates moderate-to-good reproducibility for LV global longitudinal strain.
Why the study?
Fetal speckle tracking echocardiography has not been integrated into routine pregnancy care, partly due to concerns regarding inconsistent reproducibility.
Cohort (n=124)
Blinded to previous analyses and other observer's outcomes
No
Effect estimate: ICC 0.651 (95% CI 0.582-0.712)
p-value: p=0.435 for paired t-test
Fetal speckle tracking echocardiography shows variable reproducibility, performing better in the left ventricle and before 32 weeks gestation, indicating a need for further refinement before routine clinical use.
Variable reproducibility warrants caution against routine clinical use of fetal speckle tracking; leaves open protocol standardization in prospective studies.
Fetal speckle tracking echocardiography is an ultrasound-based technique used to assess myocardial velocity and deformation of the fetal heart. Despite its potential, the method has not yet been integrated into routine pregnancy care, partly due to concerns about inconsistent reproducibility. This study aimed to evaluate the intra- and inter-observer reproducibility of global longitudinal strain measurements derived from a fixed fetal cardiac cycle, using speckle tracking echocardiography. Healthy women with singleton pregnancies were enrolled during the second trimester. From enrolment until delivery, four-chamber view clips of the fetal heart were acquired every four weeks. For intra-observer reproducibility, a single observer analyzed the same heart cycle in one DICOM clip twice for global longitudinal strain in the left and right ventricles, with a minimum interval of two weeks between assessments in a blinded manner. For inter-observer reproducibility, two independent observers analyzed the same cardiac cycle within one DICOM clip. A total of 124 women were included, yielding 632 ultrasound clips. Intra-observer reproducibility was poor to moderate for global longitudinal strain for the right and left ventricles. Inter-observer reproducibility demonstrated moderate to good reproducibility for global longitudinal strain in both ventricles. The reproducibility was generally higher in the left ventricle than in the right, and the highest reproducibility was observed before 32 weeks of gestation. In conclusion, speckle tracking echocardiography during pregnancy showed variable reproducibility of strain analysis when performed on a fixed cardiac cycle, with more consistent results in the left ventricle. These findings support the potential utility of fetal speckle tracking echocardiography, while highlighting the need for further refinement of reproducibility before clinical implementation.
No takes yet. Share an insight, caveat, or question.
Meireson et al. (2026) conducted a cohort in Healthy singleton pregnancies (n=124). Fetal speckle tracking echocardiography (fSTE) was evaluated on Intra-observer reproducibility of left ventricular global longitudinal strain (ICC 0.651, 95% CI 0.582-0.712, p=0.435 for paired t-test). Fetal speckle tracking echocardiography using a fixed cardiac cycle demonstrated moderate intra-observer reproducibility for left ventricular global longitudinal strain (ICC 0.651) and good inter-observer reproducibility (ICC 0.871).