Why the study?
This study aimed to evaluate the reproducibility and agreement of fetal cardiac shape and deformation measurements using FetalHQ software.
Does speckle tracking analysis using FetalHQ software provide reproducible measurements of cardiac shape and deformation in normal fetuses?
Does speckle tracking analysis using FetalHQ software provide reproducible measurements of cardiac shape and deformation in normal fetuses?
Speckle tracking analysis of the fetal heart provides reliable estimations of global and LV shape and deformation, but lower correlation for the RV.
Reproducibility supports FetalHQ research use; leaves open validation before clinical adoption in abnormal fetuses.
Introduction: This study aimed to evaluate reproducibility and agreement of fetal cardiac shape and deformation using FetalHQ. Methods: Fifty normal fetuses at 20–38 weeks of gestation were evaluated. Two operators independently selected an optimal cardiac cycle using FetalHQ®™software for speckle tracking analysis. Intra- and interobserver correlation coefficient and limits of agreement for cardiac shape and deformation were estimated. Results: Global cardiac markers: high correlation (r = 0.98) and agreement (mean difference, standard deviation [MD, SD] 5.07, 75.8) for ventricular area; moderate correlation (r = 0.78) and agreement (MD, SD: 0.016, 0.08) for global sphericity index (SI) and for left ventricle (LV) global strain (r = 0.65; MD, SD: −4.48, 11.9); and low but still significant correlation (r = 0.58) and agreement (MD, SD: −3.77, 12.27) for right ventricle (RV) global strain. For individual ventricular parameters: high correlation for LV ([median r; range] 0.98; 0.93–0.99) and RV (r = 0.98; 0.97–1.0) SI, and for LV (r = 0.92: 0.56–0.99) and RV (r = 0.96; 0.67–0.99) end diastolic diameters; moderate correlation for LV fractional shortening (r = 0.53; 0.87–0.98); and no significant correlation for RV fractional shortening (r = 0.36; 0.32–0.97). Inter- and intraobserver correlation and agreement were similar for all evaluated parameters. Conclusion: Speckle tracking analysis of the fetal heart provides reliable estimations of global and LV shape and deformation. Low correlation in the RV can be related to anatomical structures such as the moderator band.
No takes yet. Share an insight, caveat, or question.
Huntley et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: