Measures gamma and neutron detection efficiency in a novel spectrometer, indicating potential improvements in spectral analysis.
Measurements of the characteristics of a Compton suppressed spectrometer consisting of 9×CeBr3–NaI(Tl) phoswich detectors surrounded by 4×CsI(Tl) scintillation detectors intended for suppressing the Compton component of the spectrum have been carried out. Measurements were performed using gamma and neutron radiation sources (60Co, 152Eu, 239Pu/9Be). Key parameters of the spectrometer have been determined: the suppression factor of the Compton part of the γ spectrum, as well as the dependences of the neutron detection efficiency on their energy at various detection threshold values.
No takes yet. Share an insight, caveat, or question.
Povolotskiy et al. (2026) studied this question.