Answers are grounded in this paper and cite their sources.
Fast and accurate detection and analysis of light spectrum emerge as an important tool in several scientific fields and applications, hence demanding the design of precise, reliable, and fast instruments. In this article, an integrated spectrometer is presented based on two cascaded ring resonators covered with a phase change material. The Vernier architecture is implemented by these segments of phase change materials, allowing for a digital scan of wavelengths. Resolution smaller than 0.1 nm can be achieved, together with a footprint of ≈0.03 mm 2 and a bandwidth of the order of tens of nanometers.
No takes yet. Share an insight, caveat, or question.
Coppola et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: