We compared scatter correction (SC) in single-photon emission computed tomography (SPECT) images using effective scatter source estimation (ESSE) and the triple-energy window (TEW) method. We acquired 99mTc and 123I images of brain, myocardial, and performance phantoms containing rods with different diameters. We assessed contrast ratios (CRs) and ROI-based noise metrics (coefficient of variation, signal-to-noise ratio, and contrast-to-noise ratio [CNR] ). Under 99mTc, ESSE yielded higher CRs than TEW across all phantoms (mean difference 0.04, range 0.01–0.05) and produced the highest CNR in the myocardial phantom, improving the conspicuousness of the simulated defect. Under 123I, CR differences between ESSE and TEW were small and inconsistent (performance phantom: −0.04 to 0.06; brain phantom: −0.01 to 0.00). A Monte Carlo simulation (point source in air) showed substantial photopeak window penetration for cardiac high-resolution collimators (40.0%) but low penetration for medium-energy general-purpose collimators (5.1%), supporting photopeak contamination as a contributor to the 123I findings and potentially attenuating the apparent advantage of model-based SC that does not explicitly account for penetration photons. These findings suggest that SC selection should consider the radionuclide and imaging target and that ESSE might be a reasonable option for 99mTc myocardial imaging under the settings examined.