Key result
Partial circulatory support with the Synergy Pocket Micro-pump significantly improved hemodynamics at an average of 9.5 weeks, increasing cardiac index from 2.0 to 2.8 l/min/m2 (P<0.001).
Observational (n=27)
OBJECTIVE: Full mechanical support with a left-ventricular assist device (LVAD) is generally limited to end-stage heart-failure patients. We have been studying the safety and efficacy of the partial support Synergy® Pocket Micro-pump as bridge-to-transplant in a less-sick group of patients as a prelude to a study of its use for destination therapy. METHODS: The CircuLite® Synergy device is implanted via a small right-sided thoracotomy with an inflow cannula in the left atrium and an outflow graft connected to the right subclavian artery without the use of extracorporeal circulation. The micro-pump is the size of an AA battery, sits in the 'pacemaker' pocket subcutaneously in the right clavicular groove and pumps up to 3.0 l min⁻¹ from the left atrium to the right subclavian artery. RESULTS: The device has been implanted in 27 patients awaiting cardiac transplant (22 males), age 54.8 ± 10.0 years with ejection fraction (EF) 21 ± 6%, mean arterial pressure 73.5 ± 8.8 mm Hg, pulmonary capillary wedge pressure (PCWP) 27.5 ± 7.8 mm Hg and cardiac index (CI) 2.0 ± 0.4 l min⁻¹ m⁻². Duration of support has ranged from 6 to 281 days. Right-heart catheterization showed significant hemodynamic improvement in the short- and intermediate term after implant, with increases in CI from 2.0 ± 0.4 to 2.8 ± 0.6 l min⁻¹ m⁻² (p < 0.001) and reductions in PCWP from 28 ± 6 to 18 ± 7 mm Hg (p = 0.002) at an average of 9.5 ± 5.5 weeks. CONCLUSIONS: The Synergy device provides partial hemodynamic support and its use is associated with significantly improved hemodynamics, thus appearing to interrupt and partially reverse the progressive hemodynamic deterioration typical of end-stage heart failure. Ongoing efforts are aimed at demonstrating additional clinical benefits and continuing to further improve the risk/benefit ratio.
No takes yet. Share an insight, caveat, or question.
Meyns et al. (2010) conducted an observational in New York Heart Association Class IIIB and Early Class IV heart failure (n=27). CircuLite Synergy Pocket Micro-pump vs. Baseline (pre-implantation) was evaluated on Hemodynamic improvement (cardiac index) (p=< 0.001). Partial circulatory support with the Synergy Pocket Micro-pump significantly improved hemodynamics at an average of 9.5 weeks, increasing cardiac index from 2.0 to 2.8 l/min/m2 (P<0.001).
Synapse has enriched 5 related papers. Consider them for comparative context: