Key result
Cuff-less blood pressure estimation using ECG and PPG signals with a random forest regression model achieved mean errors of -0.04 ± 6.11 mmHg for SBP and 0.11 ± 3.62 mmHg for DBP against invasive BP.
Why the study?
Intra-arterial BP monitoring is invasive and cuff-based devices lack reliability and comfort for long-term monitoring during arrhythmias, leaving a need for continuous, cuff-less methods.
Does continuous blood pressure estimation using ECG and PPG signals accurately measure blood pressure in patients with arrhythmias compared to invasive monitoring?
Observational (n=35)
Does continuous blood pressure estimation using ECG and PPG signals accurately measure blood pressure in patients with arrhythmias compared to invasive monitoring?
A machine learning model using ECG and PPG signals can accurately estimate continuous, beat-to-beat blood pressure in patients with arrhythmias, meeting established medical device standards.
ECG-PPG BP estimation warrants caution in arrhythmias; leaves open prospective validation before clinical adoption.
OBJECTIVE: Continuous blood pressure (BP) provides valuable information for the disease management of patients with arrhythmias. The traditional intra-arterial method is too invasive for routine healthcare settings, whereas cuff-based devices are inferior in reliability and comfortable for long-term BP monitoring during arrhythmias. The study aimed to investigate an indirect method for continuous and cuff-less BP estimation based on electrocardiogram (ECG) and photoplethysmogram (PPG) signals during arrhythmias and to test its reliability for the determination of BP using invasive BP (IBP) as reference. METHODS: Thirty-five clinically stable patients (15 with ventricular arrhythmias and 20 with supraventricular arrhythmias) who had undergone radiofrequency ablation were enrolled in this study. Their ECG, PPG, and femoral arterial IBP signals were simultaneously recorded with a multi-parameter monitoring system. Fifteen features that have the potential ability in indicating beat-to-beat BP changes during arrhythmias were extracted from the ECG and PPG signals. Four machine learning algorithms, decision tree regression (DTR), support vector machine regression (SVR), adaptive boosting regression (AdaboostR), and random forest regression (RFR), were then implemented to develop the BP models. RESULTS: The results showed that the mean value ± standard deviation of root mean square error for the estimated systolic BP (SBP), diastolic BP (DBP) with the RFR model against the reference in all patients were 5.87 ± 3.13 and 3.52 ± 1.38 mmHg, respectively, which achieved the best performance among all the models. Furthermore, the mean error ± standard deviation of error between the estimated SBP and DBP with the RFR model against the reference in all patients were -0.04 ± 6.11 and 0.11 ± 3.62 mmHg, respectively, which complied with the Association for the Advancement of Medical Instrumentation and the British Hypertension Society (Grade A) standards. CONCLUSION: The results indicated that the utilization of ECG and PPG signals has the potential to enable cuff-less and continuous BP estimation in an indirect way for patients with arrhythmias.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2020) conducted an observational in Arrhythmias (n=35). Cuff-less blood pressure estimation using ECG and PPG signals vs. Invasive blood pressure (IBP) was evaluated on Root mean square error and mean error of estimated SBP and DBP against invasive BP reference. Cuff-less blood pressure estimation using ECG and PPG signals with a random forest regression model achieved mean errors of -0.04 ± 6.11 mmHg for SBP and 0.11 ± 3.62 mmHg for DBP against invasive BP.