Answers are grounded in this paper and cite their sources.
Inflammatory mediators play a major role in the development and progression of acute kidney injury (AKI). Continuous renal replacement therapy (CRRT) removes these mediators from the blood using AN69-M, AN69-ST, and HF1400 filters to target low and middle-molecular weight molecules. We characterized the in vitro removal performance of each filter in a 72 hour simulated CRRT procedure. Urea clearance with AN69-M and AN69-ST remained stable (52.4 and 51.2 ml/minute, respectively) but decreased with HF1400 (47.0 ml/minute; p < 0.001). Vancomycin clearance remained stable for AN69 filters but decreased for HF1400. Interleukin (IL)-8 was removed primarily via adsorption with the AN69 filters (92.2 and 91.2 ml/minute for AN69-M and AN69-ST, respectively), but clearance was significantly lower with HF1400 (8.4 ml/minute). Tumor necrosis factor (TNF)-α clearance was higher with AN69-ST compared with AN69-M or HF1400 (10.3, 1.8, and 2.3 ml/minute, respectively). β 2 -microglobulin clearance was higher with both AN69-based filters. The hydrogel water repartition of AN69 filters was different, with a higher percentage of bound water in AN69-ST versus AN69-M (30.5% ± 0.2% and 19.3% ± 1.5%, respectively; p < 0.05). These results suggest that clearance profiles of CRRT filters differ according to their properties; further investigation is needed to translate this into clinical improvements.
No takes yet. Share an insight, caveat, or question.
Malard et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: