Key result
Ultrasonic vascular phantoms mimicking an atheromatous plaque were successfully manufactured with Young's moduli of the vascular wall and soft inclusion at 342 ± 25 kPa and 17 ± 3 kPa, respectively.
Why the study?
Increasing complexity of ultrasound signal processing algorithms makes demonstrating their added value difficult, necessitating robust validation strategies.
The developed atherosclerotic carotid bifurcation phantoms with realistic geometries and mechanical properties provide a robust tool for validating new ultrasound-based imaging technologies.
Provides vascular phantoms for validating ultrasound plaque algorithms; leaves open translation to clinical flow or elastography use.
As the complexity of ultrasound signal processing algorithms increases, it becomes more difficult to demonstrate their added value and thus robust validation strategies are required. We propose a method of manufacturing ultrasonic vascular phantoms mimicking an atheromatous plaque in an internal carotid artery bifurcation for applications in flow imaging and elastography. During the fabrication process, a soft inclusion mimicking a stenotic lipid pool was embedded within the vascular wall. Mechanical testing measured Young's moduli of the vascular wall and soft inclusion at 342 ± 25 kPa and 17 ± 3 kPa, respectively. B-mode, color Doppler, power Doppler, shear wave elastography, and strain elastography images of the different phantoms were produced to show the validity of the fabrication process. Because of their realistic geometries and mechanical properties, those phantoms may become advantageous for fluid-structure experimental modeling and validation of new ultrasound-based imaging technologies.
No takes yet. Share an insight, caveat, or question.
Chayer et al. (2019) studied Atheromatous plaque in an internal carotid artery bifurcation (simulated). Ultrasonic vascular phantoms mimicking an atheromatous plaque was evaluated on Young's moduli of the vascular wall and soft inclusion. Ultrasonic vascular phantoms mimicking an atheromatous plaque were successfully manufactured with Young's moduli of the vascular wall and soft inclusion at 342 ± 25 kPa and 17 ± 3 kPa, respectively.