This is the authors' abstract. We don't add key points for this paper.
A simple, self-calibrating, rotating-waveplate polarimeter is largely insensitive to light intensity fluctuations and is shown to be useful for determining the Stokes parameters of light. This study shows how to minimize the in situ self-calibration time, the measurement time, and the measurement uncertainty. The suggested methods are applied to measurements of spatial variations in the linear and circular polarizations of laser light passing through glass plates with a laser intensity dependent birefringence. These are crucial measurements for the ACME electron electric dipole measurements, requiring accuracies in circular and linear polarization fraction of about 0.1% and 0.4%, with laser intensities up to 100 mW/mm 2 incident into the polarimeter.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Wirthl et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: