Why the study?
Does high-resolution 7-Tesla cardiac MRI accurately quantify myocardial mass in a mouse model compared to ex vivo measurements?
Does high-resolution 7-Tesla cardiac MRI accurately quantify myocardial mass in a mouse model compared to ex vivo measurements?
High-resolution 7-Tesla cardiac MRI provides accurate and reproducible noninvasive quantification of left ventricular mass and function in mice.
Answers are grounded in this paper and cite their sources.
Supports noninvasive LV mass quantification in mice; extends high-field MRI validation for preclinical studies but should not guide clinical practice.
The purpose of this work was to develop high-resolution cardiac magnetic resonance imaging techniques for the in vivo mouse model for quantification of myocardial function and mass. Eight male mice were investigated on a 7-Tesla MRI scanner. High-quality images in multiple short axis slices (in-plane resolution 117 microm2, slice thickness 1 mm) were acquired with an ECG-gated cine sequence. Left ventricular end-diastolic and end-systolic volumes and mass were calculated from segmented slice volumes. There was precise agreement of left ventricular mass determined ex vivo and by MRI. Intraobserver (5%) and interobserver (5%) variability of in vivo MR measurements were low.
No takes yet. Share an insight, caveat, or question.
Ruff et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: