Why the study?
Long-term data on combining third-generation S-ICD with modern software, programming strategies, and the intermuscular two-incision technique in arrhythmogenic cardiomyopathy variants were lacking.
What are the long-term outcomes and complication rates of third-generation S-ICD implanted with the IM two-incision technique in patients with arrhythmogenic cardiomyopathy?
What are the long-term outcomes and complication rates of third-generation S-ICD implanted with the IM two-incision technique in patients with arrhythmogenic cardiomyopathy?
Third-generation S-ICD with IM two-incision technique in ACM patients is associated with a low risk of cardiac oversensing, but myopotential oversensing during effort remains a cause of inappropriate shocks.
Modern S-ICD techniques may be feasible in ACM; leaves open need for randomized comparisons versus transvenous ICDs.
BACKGROUND: Long-term data on the potential advantages of combining the third-generation subcutaneous implantable cardioverter defibrillator (S-ICD) with modern software upgrade including the "SMART Pass", modern programming strategies and the intermuscular (IM) two-incision implantation technique in arrhythmogenic cardiomyopathy (ACM) with different phenotypic variants are lacking. In this study we evaluated the long-term outcome of patients with ACM who underwent third-generation S-ICD (Emblem, Boston Scientific) and IM two-incision technique. METHODS: The study population included 23 consecutive patients [70% male, median age 31 (24-46) years] diagnosed with ACM with different phenotypic variants who received third-generation S-ICD implantation with the IM two-incision technique. RESULTS: During a median follow-up of 45.5 months [16-65], 4 patients (17.4%) received a at least one inappropriate shock (IS), with median annual event rate of 4.5%. Extra-cardiac oversensing (myopotential) during effort represented the only cause of IS. No IS due to T-wave oversensing (TWOS) were recorded. Only one patient (4.3%) experienced device-related complication consisting of premature cell battery depletion requiring device replacement. No device explantation because of need for anti-tachycardia pacing or ineffective therapy occurred. There was no significant difference between patients who did and did not experienced IS with regard to baseline clinical, ECG and technical characteristics. Five patients (21.7%) received appropriate shocks on ventricular arrythmias. CONCLUSIONS: According to our findings, although the third-generation S-ICD implanted with the IM two-incision technique appears to be associated with a low risk of complications and IS due to cardiac oversensing, the risk of IS due to myopotential mainly during effort should be considered.