Measurements of the counting rates (∼103 to 104 counts/sec) of continuous-channel electron multipliers mounted in an electrostatic analyzer responding to a monoenergetic beam of electrons, while operating in a vacuum chamber at a pressure ∼3×10−6 Torr attained with an oil diffusion pump, display a degradation of their gain (fatigue) which is proportional to the accumulated counts. The useful lifetime of these devices when employed with fixed-threshold pulse amplifiers is defined here as the accumulated counts until gain degradation has produced a reduction of the counting rates to 15% of the initial responses at an operating bias voltage of 4000 V and constant stimuli. The lifetimes of these particle detectors in this laboratory environment are ∼1010 counts or, for example, an average counting rate of 300 counts/sec for one year. Comparison of this laboratory lifetime with the responses of similar instrumentation which has been flown on the earth satellites OGO's 3, 4, and 5 and IMP 4 demonstrates that the expected lifetimes for these electron multipliers in a spaceflight environment are several years. Efficiencies of an electron multiplier for counting monoenergetic electrons over an energy range ∼60 to 50 000 eV are also presented.