Ventricular pacing from the substernal space is feasible, providing proof-of-concept for an extravascular ICD configuration with pacing capabilities.
Substernal pacing proved feasible intraoperatively; leaves open whether extravascular ICDs can deliver reliable pacing in ambulatory patients.
INTRODUCTION: The objective of this study was to assess feasibility of ventricular pacing and thresholds from within the substernal space to examine a new extravascular ICD configuration with pacing capabilities. METHODS: In patients undergoing midline sternotomy, a duodecapolar diagnostic pacing catheter was positioned in the substernal space anterior to the pericardium, and a cutaneous patch in left lateral position. Different unipolar and bipolar pacing configurations were assessed. Strength-duration curves were performed to identify the optimal output, starting at 25 mA with a pulse width of 10 milliseconds. RESULTS: Eight patients with mean age 69 ± 9 years were included. In 5, ventricular capture was achieved in ≥1 configuration. The mean bipolar pacing thresholds at PW 10, 5, 3, 1 milliseconds were 12.4 ± 3.7 mA (5 patients), 13.3 ± 5.8 mA (3 patients), 18.3 ± 5.7 mA (3 patients), and 25 ± 0 mA (2 patients), respectively. The 60-mm electrode spacing was the most successful bipolar configuration. Unipolar pacing was successful in 3 out of 4 patients with mean thresholds of 10 ± 0 mA at 10 milliseconds (3 patients), 15 ± 0 mA at 5 milliseconds (3 patients), 16.7 ± 2.9 mA at 3 milliseconds (3 patients), and 20 ± 7.1 mA at 1 milliseconds (2 patients). CONCLUSION: Ventricular pacing from the substernal space in patients with midline sternotomy is feasible. Closed sternum studies are needed to determine pacing thresholds more accurately.
No takes yet. Share an insight, caveat, or question.
Brouwer et al. (2017) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: