Key result
Transpulmonary thermodilution tracks pulmonary artery thermodilution for cardiac output with ~27% error in severe HF.
Why the study?
Transpulmonary thermodilution has been adopted as an alternative to pulmonary artery thermodilution for cardiac output measurement, but no data had been published on the comparability of the two methods in patients with severely reduced left ventricular function.
Does aortic transpulmonary thermodilution provide comparable and valid cardiac output measurements compared to pulmonary artery thermodilution in patients with severe heart failure (LVEF <35%)?
Observational (n=29)
No
Does aortic transpulmonary thermodilution provide comparable and valid cardiac output measurements compared to pulmonary artery thermodilution in patients with severe heart failure (LVEF <35%)?
Answers are grounded in this paper and cite their sources.
Aortic transpulmonary thermodilution provides valid cardiac output measurements with acceptable bias and percentage error compared to pulmonary artery catheter thermodilution in patients with severe HFrEF (LVEF <35%).
Objective Hemodynamic monitoring with the pulmonary artery catheter is frequently used in the management of severe heart failure. For measurement of cardiac output (CO), transpulmonary thermodilution (TPTD) has recently been adopted into clinical practice as an alternative to pulmonary artery thermodilution. However, no data have been published on the comparability of the two methods for patients with severely reduced left ventricular function. Our objective was to evaluate the correlation between these two methods of CO determination in patients with severe left ventricular dysfunction. Design Prospective observational clinical study. Setting Cardiological intermediate care unit and medical intensive care unit of a university hospital. Patients Twenty-nine patients with left ventricular ejection fraction <35% and symptoms of heart failure (New York Heart Association class III-IV). Intervention None. Measurements and main results The two methods of intermittent CO measurement were compared by simultaneously recording the results of pulmonary artery thermodilution and TPTD after injection of a cold saline bolus. Measurements were performed when clinically necessary. A total of 325 data pairs were analyzed. Mean CO of both methods was 4.4 L/min with a bias of 0.45 L/min (2 SD 1.20 L/min), resulting in a percentage error of 27.3%. Conclusion In patients with severely impaired left ventricular function, measurement of CO by TPTD provides valid results.
No takes yet. Share an insight, caveat, or question.
Friesecke et al. (2008) conducted an observational in Severe heart failure with severely reduced left ventricular function (n=29). Aortic transpulmonary thermodilution (TPTD) after cold saline bolus injection vs. Pulmonary artery thermodilution was evaluated on Cardiac output (CO) correlation, bias, and percentage error between transpulmonary thermodilution and pulmonary artery thermodilution (Bias 0.45 L/min (2 SD 1.20 L/min), percentage error 27.3%). Transpulmonary thermodilution correlated with pulmonary artery thermodilution for cardiac output in severe heart failure with a bias of 0.45 L/min and percentage error of 27.3%.