Purpose To evaluate the clinical feasibility of a modified Goldmann applanation system capable of estimating true intraocular pressure and corneal elasticity in vivo . Methods A dual-applanation prism was developed, allowing two sequential IOP readings under identical optical conditions. A mathematical model was applied to compute true intraocular pressure and the corneal Young’s modulus measured applanation forces, corneal thickness and curvature. Fifty-three healthy eyes were examined. Model-derived true intraocular pressure values were compared with standard GAT, Pascal dynamic contour tonometry and corneal Young’s modulus. Results Mean IOPT was 15.72 ± 2.44 mmHg, significantly higher than GAT measurements (14.86 ± 2.71 mmHg; mean difference +0.86 ± 1.82 mmHg, p = 0.0012). IOPT showed strong agreement with Pascal DCT (15.60 ± 2.35 mmHg), with minimal bias (+0.15 ± 1.63 mmHg, p = 0.499) and high correlation (r = 0.93, p < 0.0001). Measurement variance was lower for IOPT than for GAT, indicating improved precision. The mean in vivo corneal Young’s modulus was 0.18 ± 0.04 MPa (range 0.057 MPa to 0.296 MPa.), consistent with physiological values. Young’s modulus showed significant age-related variation (p < 0.001) and a moderate inverse correlation with IOPT (r = −0.53, p < 0.001), while no significant sex-related differences were observed. Conclusion This dual-applanation system enables real-time in vivo estimation of both true intraocular pressure and corneal elasticity. The model provides a physically grounded correction to Goldmann tonometry and introduces a new clinically accessible method to assess corneal biomechanics.
No takes yet. Share an insight, caveat, or question.
Caruso et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: