Why the study?
Does the Architect STAT chemiluminescent microparticle immunoassay provide comparable precision and accuracy to the AxSYM MEIA technology for determining Troponin I in patients with myocardial damage?
Does the Architect STAT chemiluminescent microparticle immunoassay provide comparable precision and accuracy to the AxSYM MEIA technology for determining Troponin I in patients with myocardial damage?
The Architect chemiluminescent troponin I assay demonstrates good analytical performance and correlates highly with the established AxSYM system.
Answers are grounded in this paper and cite their sources.
Architect assay may be interchangeable with AxSYM for troponin I in labs; leaves open outcome validation before routine clinical adoption.
Introduction: The aim of this study was determination of troponin I at serum using Architect (Abbott) and AxSYM System (Abbott). Troponin is regulatory subunit of the troponin complex associate with actin filament within muscle cells and it is a marker for diagnosis of myocardial damage.Methods: We used Architect STAT chemiluminescent microparticle immunoassay (CMIA) and AxSYM microparticle Enzyme Immunoassay (MEIA), techniques for quantitative determination of cardiac TnI in human serum or plasma. At our study we have proved precision, reproducibility and accuracy from both methods. The investigation included patients (n=119) who have myocardial infarction or ischemic heart damage and were treated at cardiology, emergency, internal medicine and neurology unit in Clinical Center University in Sarajevo.Results: The precision for three controls using Architect STAT TnI asssay technology were 3.6 – 5.2 % and reproducibility was 3.7 to 5.6 %. The AxSYM STAT TnI has precision for three controls 4.3–6.6 % and reproducibilitywas from 4.8 to 7.8 %. We have got very good correlation between Architect and AxSYM technology r = 0.999 in the investigation of troponin I in serum.Conclusions: We can conclude that chemiluminescent troponin assay I (Architect) showed good analytical performance and gave new possibility at troponin I determination.
No takes yet. Share an insight, caveat, or question.
Nafija Serdarević (2011) studied this question.