Why the study?
The study aimed to develop a novel triple-bladder cuff method for accurate, automated estimation of SBP and DBP based on pulse wave dynamics, and validate its reliability in animal experiments.
Does a novel triple-bladder cuff method accurately estimate blood pressure compared to intra-arterial measurement in an animal model?
Does a novel triple-bladder cuff method accurately estimate blood pressure compared to intra-arterial measurement in an animal model?
A novel triple-bladder cuff method accurately estimates systolic and diastolic blood pressure in an animal model, showing potential for improved automated blood pressure monitoring.
Answers are grounded in this paper and cite their sources.
Animal validation of triple-bladder cuff supports further development; leaves open human efficacy testing before any clinical consideration.
Objective:The objective of this study was to develop a novel triple-bladder cuff method for accurate and automated estimation of systolic (SBP) and diastolic (DBP) blood pressure and validate its reliability in animal experiments.Methods:The cuff is composed of three bladders each measured one-third the width of a conventional BP cuff, which are designed to measure oscillatory pulsation at the proximal, middle, and distal segments of the upper arm. This structure allows evaluation of the pulse wave propagation in the brachial artery under the cuff. SBP is estimated (SBPe) by detecting resumption of systolic arterial flow based on statistical similarity in oscillatory pulse traces between the proximal and distal segments. DBP is estimated (DBPe) based on the relation between pulse wave velocity and transmural pressure at diastole in the brachial artery. In 7 anesthetized goats, we compared SBPeand DBPeto reference SBP and DBP, respectively, measured by an intra-arterial catheter. BP was perturbed by infusing nitroprusside or noradrenaline.Results:SBP correlated strongly with SBPein each animal [mean coefficient of determination (R2) = 0.98 ± 0.01]. Mean ± standard deviation of errors between SBP and SBPewas 0.0 ± 4.9 mmHg. DBP correlated strongly with DBPein each animal (R2= 0.96 ± 0.03). Mean ± standard deviation of errors between DBP and DBPewas 0.0 ± 6.3 mmHg.Conclusion:This method estimates SBP and DBP with acceptable accuracy.Significance:Accurate and automated BP estimation by this method may potentially optimize antihypertensive treatment in patients with hypertension.
No takes yet. Share an insight, caveat, or question.
Uemura et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: