A procedure has been developed to separate and quantitate phospholipids, including phosphatidylinositol and phosphatidylserine, from animal tissues by means of the Iatroscan TLC/FID technique. The method is based on the use of 0.01 M oxalic acid impregnated Chromarods-SII and stepwise resolution of the phospholipids in the presence of 1,2-dipalmitoyl-sn-glycero-3-phospho (N,N-dimethylethanolamine) as internal standard. To remove the neutral lipids, the rods are initially developed in a non-polar solvent mixture followed by partial scanning. Next, the rods are impregnated with oxalic acid, developed twice in CHCl3/CH3OH/CH3COOH/HCOOH/H2O (80:35:2:1:3, v/v/v/v/v) and partially scanned for measuring lysophosphatidylcholine, sphingomyelin and phosphatidylcholine. The subsequent step involves double development in CHCl3/CH3OH/30% NH4OH (60:35:0.9, v/v/v) to resolve cardiolipin, internal standard, phosphatidylethanolamine, phosphatidylinositol, phosphatidylserine and phosphatidic acid. For each phospholipid a linear calibration curve with a highly significant correlation coefficient was obtained. However, the calibration lines extrapolated to negative intercepts on the ordinate, indicating declining sensitivity at low phospholipid loads.
No takes yet. Share an insight, caveat, or question.
Schrijver et al. (1991) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: