Answers are grounded in this paper and cite their sources.
Directigen FLU-A, an enzyme immunoassay membrane test, was compared prospectively to isolation in cell culture and direct immunofluorescence (IF) for the detection of influenza A virus. One hundred ninety specimens were evaluated by Directigen FLU-A and cell culture; 184 of these specimens were also tested by direct IF. The sensitivity of Directigen FLU-A compared to isolation in cell culture and direct IF was 100%. The specificities of Directigen FLU-A compared to isolation and direct IF were identical, 91.6%. Fourteen specimens that were positive by Directigen FLU-A did not yield virus in culture; two of the specimens, however, were positive by direct IF, and four other specimens were not specimens of choice for the test. A positive Directigen result had positive predictive values of 62.6 and 75.0% compared to isolation and direct IF, respectively; a positive Directigen result with an intensity reading of 2+ or greater, however, had positive predictive values of 85 and 100% compared to isolation and direct IF, respectively. In all comparisons, the negative predictive value was 100%. There was no evidence that cross-reactivity occurred with non-influenza A antigens. Directigen FLU-A should serve as a convenient screening test for influenza A and as a rapid test supported by isolation in cell culture during an influenza outbreak.
No takes yet. Share an insight, caveat, or question.
Waner et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: