Answers are grounded in this paper and cite their sources.
BACKGROUND AND OBJECTIVES: The high-power Thulium fiber laser has previously been shown to rapidly vaporize and coagulate soft urinary tissues (e.g., prostate). This is the first preliminary study of a high-power Thulium fiber laser for fragmentation of urinary stones. STUDY DESIGN/MATERIALS AND METHODS: A continuous-wave, high-power Thulium fiber laser operating at a wavelength of 1.94 microm, was modulated to operate in pulsed mode with an output pulse energy of 1 J through a 300-microm-core silica fiber at a 20 milliseconds pulse length and repetition rate of 10 Hz. The fragmentation time to reduce uric acid (UA) (n = 13) and calcium oxalate monohydrate (COM) (n = 6) stones into particles < 2 mm was measured. RESULTS: Mean initial mass of the UA and COM stones measured 860+/-211 and 763 +/- 204 mg. Fragmentation rates measured 388 +/- 49 and 25 +/- 2 mg/minute. Average time needed to fragment the UA and COM stones into particles < 2 mm was 2.25 +/- 0.63 and 30.7 +/- 8.4 minutes, respectively. CONCLUSIONS: The high-power Thulium fiber laser, when operated in pulsed mode, is capable of fragmenting both soft (UA) and hard (COM) urinary stones. The Thulium fiber laser may be useful as a single laser system for use in multiple soft and hard tissue laser ablation applications in urology.
No takes yet. Share an insight, caveat, or question.
Nathaniel M. Fried (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: