This is the authors' abstract. We don't add key points for this paper.
This simplified determination of serotonin in human plasma involves "high-performance" liquid chromatography with electrochemical detection. Deproteinated samples of platelet-poor plasma are injected directly into the chromatograph without preliminary chromatographic purification. The limit of quantitative detection of serotonin in plasma is 5 nmol/L (equivalent to 250 fmol injected). Intra- and inter-assay CVs ranged from 5.6% to 7.7% and 3.5% to 14.9%, respectively, for concentrations of serotonin from 0.02 to 10.0 mumol/L. Results agreed well with those by spectrofluorometry, both for plasma standards (r = 0.985, range 100 nmol/L to 100 mumol/L) and for pathological samples obtained from four carcinoid syndrome patients (r = 0.964). The method is also applicable to the concurrent measurement of 5-hydroxytryptophan and 5-hydroxyindoleacetic acid.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Tagari et al. (1984) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: