This is the authors' abstract. We don't add key points for this paper.
We demonstrate the simultaneous estimation of signal frequency and amplitude by a single quantum sensor in a single experimental shot. Sweeping the qubit splitting linearly across a span of frequencies induces a non-adiabatic Landau-Zener transition as the qubit crosses resonance. The signal frequency determines the time of the transition, and the amplitude its extent. Continuous weak measurement of this unitary evolution informs a parameter estimator retrieving precision measurements of frequency and amplitude. Implemented on radiofrequency-dressed ultracold atoms read out by a Faraday spin-light interface, we sense a magnetic signal with 20~pT precision in amplitude, and near-transform-limited precision in frequency, in a single 300~ms sweep from $7$ to 13~kHz. The protocol realizes a swept-sine quantum spectrum analyzer, potentially sensing hundreds or thousands of channels with a single quantum sensor.