CdTe is a promising material for the detection of γ- and X-rays as 1 mm thick CdTe sensors offer an absorption efficiency that is higher than 50 {%} for photon energies up to 120 keV. Therefore 1mm thick CdTe from Acrorad has been flip-chipped as the sensor material on a Timepix readout-electronics ASIC at the Freiburg Materials Research Center (FMF). The transport properties of the CdTe material are investigated by determining the μτ-product of the charge carriers by illuminating the non-collecting electrode with alpha particles. The method commonly used to determine the μτ-product for non-pixelated devices was adapted for the Medipix pixelated readout-electronics. The position of the alpha particles on the pixel matrix provides spatial information, which was used to create a mapping of the μτ-product. The energy resolutions and the positions of the noise edge of two CdTe-Timepix assemblies, one with a pixel pitch of 110 x 110 μm 2 and another one with a pixel pitch of 55 x 55 μm 2 were investigated by performing a threshold scan of a 241 Am source. A comparison of the energy spectra obtained with the two assemblies with different pixel pitches confirms the strong dependence of charge sharing on the energy spectra, as expected.
No takes yet. Share an insight, caveat, or question.
Greiffenberg et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: