Why the study?
Does implantation of a single pass lead VDD pacemaker provide reliable atrial sensing and ventricular pacing in patients with second- to third-degree AV block?
Does implantation of a single pass lead VDD pacemaker provide reliable atrial sensing and ventricular pacing in patients with second- to third-degree AV block?
Single pass lead VDD pacing is reliable and easy to manage, providing dependable atrial sensing and ventricular pacing up to 6 months.
Answers are grounded in this paper and cite their sources.
Single-pass VDD pacing may be feasible in AV block; leaves open confirmation of reliability in prospective randomized trials.
UNLABELLED: Twenty-five patients with second- to third-degree AV block and normal sinus function (16 males, mean age 60 +/- 18; range 15-78 years) underwent implantation of VVD pacemakers (THERA VDD, Medtronic, Inc.) with a single pass (SP) lead. RESULTS: During implantation the mean amplitude of the atrial (A) signal was 3.9 +/- 1.4 mV (range 2.0-7.8 mV). Stable, acceptable A-signals during implantation were usually observed in the mid- or lower part of the right atrium. The lead tip electrical parameters were not compromised in any patient in order to obtain an acceptable A-signal. To verify VDD device function, patients underwent pacing system analysis on the second day and again 1, 3, and 6 months after implantation. Acute and chronic electrical measurements in the ventricle were similar to those with regular steroid leads. During follow-up tests, stable atrial sensing (A > or = 0.7 mV) was found in all but one patient (in whom A was 0.25-0.5 mV and an intermittent loss of atrial sensing occurred). There was no difference between serial measurements of A-signal amplitudes on the second day or 1, 3, and 6 months after implantation: 1.9 +/- 1.3 mV, 1.5 +/- 0.6 mV, 1.3 +/- 0.8 mV, and 1.5 +/- 1.1 mV, respectively. The mean implantation time was 54.0 +/- 17 minutes and the mean fluoroscopy time was 3.2 +/- 1.3 minutes. CONCLUSIONS: SP lead VDD pacing is reliable and easy to manage with dependable atrial sensing and ventricular pacing.(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
Ovsyshcher et al. (1994) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: