A photoluminescent thermometer, based on the transient emission response of a europium-activated calcium sulphide phosphor, is described in detail. This optical thermometry system has special advantages for temperature measurements in microwave and RF fields and potential application in electromagnetically induced clinical hyperthermia. A laboratory system has been constructed which utilises a fibreoptic probe with external diameter 0.8 mm. A system temperature resolution of 0.1 degree C has been achieved in the range 37-47 degrees C.
No takes yet. Share an insight, caveat, or question.
Samulski et al. (1982) studied this question.