Why the study?
Does tissue Doppler echocardiography (TDE) using standard ultrasound equipment allow for the detection of cardiac structures and measurement of wall velocities in normal fetuses?
Does tissue Doppler echocardiography (TDE) using standard ultrasound equipment allow for the detection of cardiac structures and measurement of wall velocities in normal fetuses?
Tissue Doppler echocardiography can be successfully performed to assess fetal cardiac function using standard high-resolution ultrasound systems without the need for specialized hardware or software.
Answers are grounded in this paper and cite their sources.
TDE feasible for fetal assessment with standard ultrasound; leaves open clinical utility pending prospective validation.
OBJECTIVES: Tissue Doppler echocardiography (TDE) has been developed in adult cardiology, but only recently has it been applied to fetal heart studies. We implemented TDE on a high-resolution ultrasound system used for prenatal scanning by changing the Doppler settings, but without specific TDE equipment, to study cardiac tissue motion of normal fetuses in the second and third trimesters. METHODS: Using color Doppler settings optimized for low velocity motion assessment, detection rates of cardiac structures were measured in 47 fetuses. With pulsed Doppler, axial diastolic and systolic cardiac wall velocities were recorded in another 30 fetuses. RESULTS: In all fetuses TDE could detect cardiac wall motion. The right ventricular wall and the level of the atrioventricular valves were most easily seen. Axial contraction velocities of the fetal heart increased with gestational age. CONCLUSIONS: TDE can be used to study normal fetal cardiac function throughout gestation. A regular ultrasound system without additional hardware or software is suitable for screening studies using TDE.
No takes yet. Share an insight, caveat, or question.
Tutschek et al. (2003) studied this question.