Key result
IABP length mismatch causes fatal visceral ischemia and multiple organ failure in an elderly patient.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
We present a case of visceral ischemia due to an anatomic versus balloon length mismatch. An 80-year-old women underwent elective coronary artery bypass graft surgery. The direct post-operative period was complicated by low cardiac output, not responding to inotropic or vasopressor drugs. An intra-aortic balloon pump (IABP) was inserted during an explorative rethoracotomy. The first 3 days, an improvement of the hemodynamic function was seen. During this period, hepatic dysfunction, renal failure with need for hemodialysis and rising values of serum lactate were seen. A computed tomography (CT) of the chest and abdomen showed splenic infarction, ascites and signs of gastric and intestinal ischemia. The IABP was blocking the celiac trunk, the superior mesenteric artery and the renal arteries and was immediately removed. The patient died shortly after due to multiple organ failure. To prevent ischemic complications in smaller patients, we advise a follow-up CT after placement of the IABP.
No takes yet. Share an insight, caveat, or question.
Claes et al. (2014) conducted a case report in Low cardiac output post-CABG complicated by visceral ischemia (n=1). Intra-aortic balloon pump (IABP) was evaluated on Visceral ischemia and multiple organ failure. Intra-aortic balloon pump placement in an 80-year-old woman caused fatal visceral ischemia and multiple organ failure due to an anatomic versus balloon length mismatch.