Answers are grounded in this paper and cite their sources.
The local concentration of potassium ion in a single HeLa cell was observed with an ion-selective nano-pipette probe. Ion selectivity was achieved by using a polyvinyl chloride film with selected ionophores placed within the nano-pipette. Both alternating and constant bias voltages were applied to the counter electrode for the observation of local ion concentrations with a response time of less than 0.1 s. These measurements were enabled by a low-current detection system prepared specifically for this study. The difference in local potassium concentrations between inside a living HeLa cell and the surrounding solution was approximately 100 mM, while no difference in potassium ion concentration was observed between the interior of dead cells and the surrounding solution.
No takes yet. Share an insight, caveat, or question.
Takami et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: