Answers are grounded in this paper and cite their sources.
Using the National Institute of Standards and Technology high-accuracy cryogenic radiometer (HACR), we have realized a scale of absolute spectral response between 406 and 920 nm. The HACR, an electrical-substitution radiometer operating at cryogenic temperatures, achieves a combined relative standard uncertainty of 0.021%. Silicon photodiode light-trapping detectors were calibrated against the HACR with a typical relative standard uncertainty of 0.03% at nine laser wavelengths between 406 and 920 nm. Modeling of the quantum efficiency of these detectors yields their responsivity throughout this range with comparable accuracy.
No takes yet. Share an insight, caveat, or question.
Gentile et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: