BACKGROUND AND OBJECTIVE: Previous studies in laser assisted vascular welding have been limited by the lack of a reliable end point for tissue fusion. As a means of improving the reproductibility of laser assisted repairs, a system incorporating real time temperature monitoring and closed loop feedback was used. STUDY DESIGN/MATERIALS AND METHODS: The system consisted of a direct view infrared thermometer for monitoring the laser heated spot, a 1.9 microns diode laser, and a microprocessor for data acquisition and feedback control of the laser power to maintain a constant tissue temperature. Rat aortas were welded under constant surface temperature conditions. RESULTS: In vivo temperature stability of +/- 2 degrees C was achieved over a temperature range of 70-90 degrees C pertinent to welding small vessels. When welds were completed using the feedback system to maintain the tissue temperature at 80 degrees C, the acute success rate was 100% and the burst pressure was 290 +/- 70 mmHg. CONCLUSION: These studies demonstrate that the use of real time monitoring and feedback control results in improved consistency for vascular tissue welding.
No takes yet. Share an insight, caveat, or question.
Stewart et al. (1996) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: