The paper describes a technical design for a fetal heart rate monitor using scalp electrodes, demonstrating feasibility of beat-to-beat monitoring using specific hardware and software configurations.
The design of a fetal heart rate (FHR) monitor using fetal electrocardiogram (FECG) scalp electrodes is described. It is shown that the design approach followed two stages: generation of FHR pulses at R-R intervals and FHR computation. The former uses a simple hardware approach for QRS detection and R-wave enhancement, while the latter requires a software implementation in order to produce FHR traces on a beat to beat basis. The QRS detection is based on bandpass filtering using switched mode capacitor technique; the R-wave enhancement and amplitude information are achieved by differentiation followed by fullwave rectification and peak detection. An adaptive threshold together with a comparator circuit are used to generate FHR pulses at R-R intervals. Beat to beat variations of FHR traces are produced by hardware and software implementation on a Z80 microprocessor board. Results obtained by the FHR monitor are evaluated and contrasted to other commercial FHR monitors.
No takes yet. Share an insight, caveat, or question.
Reguig et al. (1996) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: