An accurate light scattering photometer capable of measuring both the angular distribution of intensity and the intensity fluctuation spectrum of laser light scattered from a wide variety of samples including solutions of polymers and other macromolecules is described. The photometer employs 18 fixed scattering angles ranging from 2.6° to 163° and is sufficiently accurate to measure a molecular radius of gyration as small as 60 Å reliably. Stray elastically scattered light normally comprises less than 1% of the light scattered from pure toluene at the smallest scattering angle and is unmeasurable at other angles. The apparatus employs optical fibers to transmit the scattered light to a single photomultiplier tube and functions automatically under microprocessor control, with sample temperature controlled to ±0.1 mK.