Answers are grounded in this paper and cite their sources.
The integration of an online aerodynamic particle sizer as a Process Analytical Technology for real-time aerodynamic particle sizing in a laboratory spray dryer is studied. It is found that the aerodynamic particle sizer can discriminate inhalable powders in terms of aerodynamic size, provided that the apparatus can regulate the aerosol particle concentration via a diluter and an adequate sampling time for statistical convergence is established (20 s in the presented case study). The precision and accuracy of the real-time data from our apparatus were found to be adequate for the formulation development of composite particles for dry powder inhalers. The measurement of primary aerodynamic size allows the decoupling of formulation and process development of the spray drying operation from the downstream processes and the device performance. This decoupling accelerates development by reducing long lead time testing to characterize Dry Powder Inhaler formulations such as inertial impaction, which is the current standard.
No takes yet. Share an insight, caveat, or question.
Lopes et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: